25-7
5.
Open the relief valve to vent standard CO
2
and make sure that there is an excess gas
flow through the T-shape connector to air.
6.
Check the realtime CO
2
value is within 6.0±0.3%
in the
[
Calibrate CO2
] menu.
25.6.3 Calibrating CO
2
For the CO
2
module, a calibration is needed every year or when the measured values have a
great deviation. For maintream CO
2
module, no calibration is needed. Calibration for the CO
2
module can be performed only when the module enters the full accuracy mode.
WARNING
z
Connect an exhaust tube to the gas outlet connector of the monitor to remove the
calibration gases to a scavenging system.
Tools required:
A steel gas cylinder with 6±0.05% CO
2
and balance gas N
2
T-shape connector
Tubing
Follow this procedure to perform a calibration:
1.
Make sure that the CO
2
module has been warmed up or started up.
2.
Check the airway for leakage and perform a leakage test as well to make sure the airway
has no leakage.
3. Select
[
Main Menu
] [
Maintenance >>
] [
User Maintenance
>>] enter the
required password
ė
[
Module Maintenance >>
] [
Maintain CO
2
>>
] [
Calibrate
CO
2
>>
].
4. In
the
[
Calibrate CO
2
] menu, select [
Zero
].
5.
After the zero calibration is finished successfully, connect the equipment as follows:
Summary of Contents for uMEC10
Page 1: ...uMEC10 uMEC12 uMEC15 Patient Monitor Operator s Manual ...
Page 2: ......
Page 24: ...1 6 FOR YOUR NOTES ...
Page 29: ...2 5 2 3 2 uMEC12 uMEC15 1 6 7 2 8 9 10 3 4 5 ...
Page 36: ...2 12 FOR YOUR NOTES ...
Page 50: ...3 14 FOR YOUR NOTES ...
Page 58: ...4 8 FOR YOUR NOTES ...
Page 72: ...6 8 FOR YOUR NOTES ...
Page 118: ...8 28 FOR YOUR NOTES ...
Page 174: ...18 8 FOR YOUR NOTES ...
Page 186: ...20 6 FOR YOUR NOTES ...
Page 192: ...21 6 FOR YOUR NOTES ...
Page 202: ...23 4 FOR YOUR NOTES ...
Page 206: ...24 4 FOR YOUR NOTES ...
Page 224: ...26 10 FOR YOUR NOTES ...
Page 248: ...B 8 FOR YOUR NOTES ...
Page 266: ...C 18 FOR YOUR NOTES ...
Page 280: ...E 6 FOR YOUR NOTES ...
Page 286: ...F 6 FOR YOUR NOTES ...
Page 287: ......
Page 288: ...PN 046 008784 00 1 0 ...