16-2
16.3 Measurement Limitations
The following factors may influence the measurement accuracy:
Leaks or internal venting of sampled gas
Mechanical shock
Cyclic pressure up to 10 kPa (100 cmH
2
O)
Other sources of interference, if any
Measurement accuracy may be affected by the breath rate and inspiration/expiration (I/E)
ratio as follow:
EtCO
2
value is within specification for breath rate
60 bpm and I/E ratio
1: 1;
EtCO
2
value is within specification for breath rate
30 bpm and I/E ratio
2: 1.
Measurement accuracy is unspecified for breath rate larger than 60 bpm.
16.4 Understanding CO
2
Display
The CO
2
parameter and waveform areas provide FiCO
2
measurement, EtCO
2
measurement,
awRR measurement, and a CO
2
waveform.
(1) CO
2
waveform
(2) Fraction of inspired CO
2
(FiCO
2
)
(3) End
tidal
CO
2
value (EtCO
2
)
(4) Airway respiration rate (awRR)
(1)
(3)
(4)
(2)
Summary of Contents for uMEC10
Page 1: ...uMEC10 uMEC12 uMEC15 Patient Monitor Operator s Manual ...
Page 2: ......
Page 24: ...1 6 FOR YOUR NOTES ...
Page 29: ...2 5 2 3 2 uMEC12 uMEC15 1 6 7 2 8 9 10 3 4 5 ...
Page 36: ...2 12 FOR YOUR NOTES ...
Page 50: ...3 14 FOR YOUR NOTES ...
Page 58: ...4 8 FOR YOUR NOTES ...
Page 72: ...6 8 FOR YOUR NOTES ...
Page 118: ...8 28 FOR YOUR NOTES ...
Page 174: ...18 8 FOR YOUR NOTES ...
Page 186: ...20 6 FOR YOUR NOTES ...
Page 192: ...21 6 FOR YOUR NOTES ...
Page 202: ...23 4 FOR YOUR NOTES ...
Page 206: ...24 4 FOR YOUR NOTES ...
Page 224: ...26 10 FOR YOUR NOTES ...
Page 248: ...B 8 FOR YOUR NOTES ...
Page 266: ...C 18 FOR YOUR NOTES ...
Page 280: ...E 6 FOR YOUR NOTES ...
Page 286: ...F 6 FOR YOUR NOTES ...
Page 287: ......
Page 288: ...PN 046 008784 00 1 0 ...