7-7
7.4.5 Setting the Reminder Tones
When the alarm volume is set to zero, or the alarm tone is reset or switched off, the patient
monitor issues a periodical reminder tone. To set the reminder tones:
1. Select
[
Main Menu
][
Maintenance >>
][
User Maintenance >>
]enter the
required password.
2. Select
[
Alarm Setup >>
] to enter the [
Alarm Setup
] menu.
3. Set
the
[
Reminder Tones
] to [
On
], [
Off
] or [
Re-alarm
]. When [
Re-alarm
] is selected,
the current physiological alarms and technical alarms marked with “
Ĝ
” will be
re-generated after the [
Reminder Interval
] if the alarm condition persists.
To set the interval between reminder tones, select [
Reminder Interval
] and toggle between
[
1min
], [
2min
] and [
3min
].
In addition, you can set the volume of alarm reminder tones. To set the volume of alarm
reminder tones, select [
Main Menu
][
Alarm Setup >>
][
Others
] or the [
Alarm Setup
]
QuickKey[
Others
]. Then, select [
Reminder Vol
] and toggle between [
High
], [
Medium
]
and [
Low
].
7.5 Understanding the Alarm Setup Menu
Select the [
Alarm Setup
] QuickKey or [
Main Menu
][
Alarm Setup >>
] to enter the
[
Alarm Setup
], where you can:
Set alarm properties for all parameters.
Change ST alarm settings.
Change arrhythmia alarm settings.
Set the threshold for some arrhythmia alarms.
Change other settings.
Summary of Contents for uMEC10
Page 1: ...uMEC10 uMEC12 uMEC15 Patient Monitor Operator s Manual ...
Page 2: ......
Page 24: ...1 6 FOR YOUR NOTES ...
Page 29: ...2 5 2 3 2 uMEC12 uMEC15 1 6 7 2 8 9 10 3 4 5 ...
Page 36: ...2 12 FOR YOUR NOTES ...
Page 50: ...3 14 FOR YOUR NOTES ...
Page 58: ...4 8 FOR YOUR NOTES ...
Page 72: ...6 8 FOR YOUR NOTES ...
Page 118: ...8 28 FOR YOUR NOTES ...
Page 174: ...18 8 FOR YOUR NOTES ...
Page 186: ...20 6 FOR YOUR NOTES ...
Page 192: ...21 6 FOR YOUR NOTES ...
Page 202: ...23 4 FOR YOUR NOTES ...
Page 206: ...24 4 FOR YOUR NOTES ...
Page 224: ...26 10 FOR YOUR NOTES ...
Page 248: ...B 8 FOR YOUR NOTES ...
Page 266: ...C 18 FOR YOUR NOTES ...
Page 280: ...E 6 FOR YOUR NOTES ...
Page 286: ...F 6 FOR YOUR NOTES ...
Page 287: ......
Page 288: ...PN 046 008784 00 1 0 ...