A-16
awRR measurement range
0 to 150 rpm
awRR measurement
precision
<60 rpm: ±1 rpm
60 to 150 rpm: ±2 rpm
Apnea time
10 s, 15 s, 20 s, 25 s, 30 s, 35 s, 40 s
Note: The response time is the sum of the rise time and the delay time.
Effect of interference gases on CO
2
measurements
Gas
Concentration (%)
Quantitive effect*
N
2
O 60
±1 mmHg
Hal 4
Sev 5
Iso 5
Enf 5
O
2
100
Des 15
±2
mmHg
*: means an extra error should be added in case of gas interference when CO
2
measurements are
performed between 0-40mmHg.
Inaccuracy specifications are affected by the breath rate and I: E change. The end-tidal gas
reading is within specification for breath rate below 15BPM and I: E ratio smaller than 1: 1
relative to the gas readings without breath.
Alarm limit
Range
Step
EtCO
2
High
(low limit + 2) to 99 mmHg
1 mmHg
EtCO
2
Low
1 to (high limit - 2) mmHg
FiCO
2
High
1 to 99 mmHg
awRR High
Adult, pediatric:
Neonate:
(low limit + 2) to 100 rpm
(low limit + 2) to 150 rpm
1 rpm
awRR Low
0 to (high limit - 2) rpm
Summary of Contents for uMEC10
Page 1: ...uMEC10 uMEC12 uMEC15 Patient Monitor Operator s Manual ...
Page 2: ......
Page 24: ...1 6 FOR YOUR NOTES ...
Page 29: ...2 5 2 3 2 uMEC12 uMEC15 1 6 7 2 8 9 10 3 4 5 ...
Page 36: ...2 12 FOR YOUR NOTES ...
Page 50: ...3 14 FOR YOUR NOTES ...
Page 58: ...4 8 FOR YOUR NOTES ...
Page 72: ...6 8 FOR YOUR NOTES ...
Page 118: ...8 28 FOR YOUR NOTES ...
Page 174: ...18 8 FOR YOUR NOTES ...
Page 186: ...20 6 FOR YOUR NOTES ...
Page 192: ...21 6 FOR YOUR NOTES ...
Page 202: ...23 4 FOR YOUR NOTES ...
Page 206: ...24 4 FOR YOUR NOTES ...
Page 224: ...26 10 FOR YOUR NOTES ...
Page 248: ...B 8 FOR YOUR NOTES ...
Page 266: ...C 18 FOR YOUR NOTES ...
Page 280: ...E 6 FOR YOUR NOTES ...
Page 286: ...F 6 FOR YOUR NOTES ...
Page 287: ......
Page 288: ...PN 046 008784 00 1 0 ...