Accutorr 7 Operator’s Manual
C - 3
C.1.3
PR
C.1.4
Temp
Name
Default Config
Saved at
Power Failure
Affected by
Defaults
Alarm On/Off
On
Yes
Yes
Alarm Priority
Med
PR High
Adult
120
Pediatric
160
Neonate
200
PR Low
Adult
50
Pediatric
75
Neonate
100
PR Source
SpO
2
Beat Volume
2
Name
Default Config
Saved at
Power Failure
Affected by
Defaults
Alarm On/Off
Off
Yes
Yes
Alarm Priority
Med
Yes
Yes
Temp High
38.0
Yes
Yes
Temp Low
35.0
Yes
Yes
Temp Type
Predictive
No
No
Temp Position
Oral/Axillary probe:
Oral for adult and
pediatric
Axillary for neonate
No
No
Rectal probe: Rectal
Summary of Contents for Accutorr 7
Page 1: ...Accutorr 7 Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 22: ...1 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 30: ...2 8 Accutorr 7 Operator s Manual Trend Screen 2 6 3 9 5 1 8 7 ...
Page 76: ...6 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 84: ...7 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 86: ...8 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 96: ...9 10 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 102: ...10 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 136: ...15 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 164: ...A 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 172: ...B 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 180: ...C 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 198: ...F 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 199: ......