D - 4
Accutorr 7 Operator’s Manual
NIBP Signal
Saturated
Yes
Low
The NIBP signal is
saturated
.
Check the patient’s
condition and
reduce the patient
motion or other
sources
.
NIBP Air Leak
Yes
Low
The NIBP airway
may leak air.
The NIBP cuff is not
properly connected,
or there is a leak in
the airway.
NIBP Equip Err
Yes
High
System error; or
pump, A/D
sampling or
pressure transducer
error; or pointer
error during
software running.
Check the patient’s
condition and NIBP
connections, or
replace the cuff.
NIBP Timeout
Yes
Low
Time is out. In
Adult/Pediatric
mode, the
measurement time
is over 120 seconds;
in neonate mode,
the time is over 90
seconds.
NIBP Cuff Type
Wrong
Yes
Low
The cuff type
applied mismatches
the patient
category.
Verify the patient
category and
replace the cuff.
NIBP Illegally
Reset
Yes
Low
An illegal reset
occurred during
NIBP measurement.
Check if the airway
is occluded.
VeniPuncture
timeout
Yes
Low
System deflates the
cuff after a certain
time.
No operation is
required.
Alarm Message
Clearable?
(Yes/No)
Alarm
Level
Cause
Solution
Summary of Contents for Accutorr 7
Page 1: ...Accutorr 7 Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 22: ...1 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 30: ...2 8 Accutorr 7 Operator s Manual Trend Screen 2 6 3 9 5 1 8 7 ...
Page 76: ...6 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 84: ...7 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 86: ...8 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 96: ...9 10 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 102: ...10 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 136: ...15 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 164: ...A 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 172: ...B 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 180: ...C 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 198: ...F 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 199: ......