Accutorr 7 Operator’s Manual
6 - 13
Alarm conditions are indicated to nurses only when:
1.
The nurse call system is enabled
2.
An alarm that meets your preset requirements occurs, and
3.
The monitor is not in the alarm paused or silence status.
If no setting is selected from [
Alarm Level
] or [
Alarm Category
], no nurse call signal will
be triggered if alarms occur.
WARNING
•
To obtain the nurse call signal, use the nurse call cable (
PN: 009-003116-00
) we
supply. Otherwise the nurse call function will not work and the monitor may
be damaged.
•
Do not rely exclusively on the nurse call system for alarm notification.
Remember that the most reliable alarm notification combines audible and
visual alarm indications with the patient’s clinical condition.
Summary of Contents for Accutorr 7
Page 1: ...Accutorr 7 Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 22: ...1 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 30: ...2 8 Accutorr 7 Operator s Manual Trend Screen 2 6 3 9 5 1 8 7 ...
Page 76: ...6 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 84: ...7 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 86: ...8 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 96: ...9 10 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 102: ...10 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 136: ...15 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 164: ...A 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 172: ...B 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 180: ...C 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 198: ...F 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 199: ......