3 - 4
Accutorr 7 Operator’s Manual
monitor; insert the CO
2
sampling line into the CO
2
adapter that is installed on the side of
the monitor.
3.3
Turning On/Off Power
3.3.1
Turning Power On
Once the monitor is installed, before beginning measurements.
1.
Check the monitor for any mechanical damage, and make sure that all external
cables, plug-ins and accessories are properly connected.
2.
Check the power supply specification is met if mains power is used. Only use a
power outlet that is properly grounded.
3.
Plug the power cord into the AC power source. If you run the monitor on battery
power, ensure that the battery is sufficiently charged.
4.
Press the power on/off switch on the monitor’s front panel
.
The monitor will perform alarm system self-test during start-up. After pressing the
power on/off button, the system sounds a beep, and the alarm lamp simultaneously
turns yellow, then red, and then turns off, with the start-up screen being shown. Then
the start-up screen disappears. The alarm system self-test succeeds. The monitor enters
the normal monitoring screen.
WARNING
•
Check that visual and auditory alarm signals are presented correctly when
the equipment is powered on. Do not use the equipment for any monitoring
procedure on a patient if you suspect it is not working properly, or if it is
mechanically damaged. Contact your service personnel or Mindray
.
NOTE
•
Carefully check if the system performs the self-test as described above.
Contact your service personnel or Mindray if the self-test is abnormal.
3.3.2
Turning off the Monitor
Before turning off the monitor:
1.
Ensure that monitoring of the patient has been completed
.
2.
Disconnect cables and sensors from the patient.
3.
Make sure to save or clear the patient monitoring data as required.
Then press and hold the power on/off switch for more than 2 seconds to turn off
monitor. There is a prompt message when the system is shutting down.
Summary of Contents for Accutorr 7
Page 1: ...Accutorr 7 Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 22: ...1 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 30: ...2 8 Accutorr 7 Operator s Manual Trend Screen 2 6 3 9 5 1 8 7 ...
Page 76: ...6 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 84: ...7 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 86: ...8 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 96: ...9 10 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 102: ...10 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 136: ...15 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 164: ...A 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 172: ...B 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 180: ...C 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 198: ...F 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 199: ......