A - 12
Accutorr 7 Operator’s Manual
Accuracy*
Full accuracy mode:
0≤CO
2
<1%:
1%≤CO
2
<5%:
5%≤CO
2
<7%:
7%≤CO
2
<12%:
12%≤CO
2
≤13%:
13%<CO
2
≤20%:
±0.1%
±0.2%
±0.3%
±0.4%
±0.5%
±0.43%+8% rel
ISO accuracy mode: Add ±0.3% ABS to the full accuracy
mode
Start-up time
20 s (typical), 90 s (maximum)
Accuracy drift
Meets the requirement for measurement accuracy within 6
hours.
Resolution
1mmHg
Sample flowrate
Connecting the Oridion sampling line: 50 ml/min
Sample flowrate tolerance
±15% or ±15 ml/min, whichever is greater.
Rise time
Measured with a Oridion sampling line: ≤250 ms @ 50 ml/
min (standard sampling line) or ≤280 ms @ 50 ml/min
(extended sampling line)
Response time
Measured with a Oridion sampling line: ≤5 s @ 50 ml/min
(standard sampling line) or ≤6.5 s @ 50 ml/min (extended
sampling line)
awRR measurement range
0 to 150 rpm
awRR accuracy
<60 rpm: ±1 rpm
60 to 150 rpm: ±2 rpm
awRR resolution
1 rpm
Alarm limit
Range
Step
EtCO
2
High
(low limit + 2) to 99 mmHg
1 mmHg
EtCO
2
Low
1 to (high limit - 2) mmHg
FiCO
2
High
1 to 99 mmHg
awRR High
Adult, pediatric: (low limit +
2) to 100 rpm
Neonate: (low limit + 2) to
150 rpm
1 rpm
awRR Low
0 to (high limit - 2) rpm
Summary of Contents for Accutorr 7
Page 1: ...Accutorr 7 Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 22: ...1 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 30: ...2 8 Accutorr 7 Operator s Manual Trend Screen 2 6 3 9 5 1 8 7 ...
Page 76: ...6 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 84: ...7 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 86: ...8 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 96: ...9 10 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 102: ...10 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 136: ...15 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 164: ...A 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 172: ...B 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 180: ...C 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 198: ...F 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 199: ......