9 - 6
Accutorr 7 Operator’s Manual
5.
Time of last measurement
6.
Time remaining to next measurement
7.
Measurement mode
9.7
Setting NIBP
Select NIBP parameter area to access the [
NIBP Setup
] menu.
9.7.1
Setting Interval
In [
NIBP Setup
] menu, you can select [
Interval
] and set to:
■
[
Manual
]: NIBP measurement is started manually.
■
[
1 min
], [
2 min
], [
2.5 min
], [
3 min
], [
5 min
], [
10 min
], [
15 min
], [
20 min
], [
30 min
],
[
1 h
], [
1.5 h
], [
2 h
]: The monitor automatically measures NIBP based on the
specified time interval.
■
[
Program 1
] and [
Program 2
]: The monitor automatically measures NIBP based on
a program configured by user.
9.7.2
Setting the Initial Cuff Inflation Pressure
You can set the initial cuff inflation pressure manually. In the [
NIBP Setup
] menu, select
[
Initial Pressure
] and then select the appropriate setting.
The initial inflation pressure range is as follows:
NOTE
•
For known hypertensive patients, you need to set the initial cuff pressure to a
higher value to reduce the measurement time.
9.7.3
Setting NIBP End Tone
The monitor can issue a reminder tone at the completion of NIBP measurement. The
NIBP End Tone is off by default. You can switch it on by accessing the [
NIBP Setup
]
menu.
Patient Category
Range (mmHg)
Default (mmHg)
Adult
80 - 280
160
Pediatric
80 - 210
140
Neonate
60 - 140
90
Summary of Contents for Accutorr 7
Page 1: ...Accutorr 7 Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 22: ...1 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 30: ...2 8 Accutorr 7 Operator s Manual Trend Screen 2 6 3 9 5 1 8 7 ...
Page 76: ...6 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 84: ...7 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 86: ...8 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 96: ...9 10 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 102: ...10 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 136: ...15 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 164: ...A 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 172: ...B 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 180: ...C 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 198: ...F 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 199: ......