A - 6
Accutorr 7 Operator’s Manual
A.7
Measurement Specifications
A.7.1
SpO
2
Masimo SpO
2
Module
Alarm limit
Range (%)
Step (%)
SpO
2
High
(low limit + 2) to 100
1
SpO
2
Low
Masimo: Desat to (high limit – 2)
Nellcor: Desat or 20 (whichever is greater) to
(high limit – 2)
Desat
0 to (high limit – 2)
Standards
Meet standards of ISO 9919, ISO 80601-2-61
SpO
2
Measurement range
1 to 100%
PI measurement range
0.02% to 20%
SpO
2
Resolution
1%
Accuracy
1
70 to 100%: ±2% (measured without motion in adult/
pediatric mode)
70 to 100%: ±3% (measured without motion in neonate
mode)
70 to 100%: ±3% (measured with motion)
1% to 69%: Not specified.
Refreshing rate
1 s
Response time
≤ 20 s (PR 75 bpm, average time 8 s, SpO
2
value rises
from 60% to 95%)
SpO
2
averaging time
2-4 s, 4-6 s, 8 s, 10 s, 12 s, 14 s, 16 s
Low perfusion conditions
Pulse amplitude: >0.02%
Light penetration: >5%
Low perfusion SpO
2
accuracy
2
±2%
Summary of Contents for Accutorr 7
Page 1: ...Accutorr 7 Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 22: ...1 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 30: ...2 8 Accutorr 7 Operator s Manual Trend Screen 2 6 3 9 5 1 8 7 ...
Page 76: ...6 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 84: ...7 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 86: ...8 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 96: ...9 10 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 102: ...10 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 136: ...15 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 164: ...A 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 172: ...B 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 180: ...C 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 198: ...F 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 199: ......