Accutorr 7 Operator’s Manual
6 - 7
NIBP-S
(SYS×0.68)
+
10 mmHg
(SYS-15) or 45
mmHg,
whichever is
greater
(SYS×0.86)
+
38 mmHg
(SYS+15) or
105 mmHg,
whichever is
smaller
Adult:
45 to 270
mmHg
Pediatric:
45 to 185
mmHg
Neonate:
35 to 115
mmHg
NIBP-D
(Dia×0.68)
+
6
mmHg
(Dia-15) or 20
mmHg,
whichever is
greater
(Dia×0.86)
+
32
mmHg
(Dia
+
15) or
80 mmHg,
whichever is
smaller
Adult:
25 to 225
mmHg
Pediatric:
25 to 150
mmHg
Neonate:
20 to 90 mmHg
NIBP-M
(Mean×0.68)
+
8 mmHg
(Mean-15) or
35 mmHg,
whichever is
greater
(Mean×0.86)
+
35 mmHg
(Mean
+
15)
or 95 mmHg,
whichever is
smaller
Adult:
30 to 245
mmHg
Pediatric:
30 to 180
mmHg
Neonate:
25 to 105
mmHg
Temp
(T
-
0.5) °C
(T-0.5) °C
(T+0.5) °C
(T+0.5) °C
Same as the
measurement
range
Parameter
Low alarm Limit
High alarm Limit
Auto-set alarm
limit Range
Adult/
Pediatric
Neonate
Adult/
Pediatric
Neonate
Summary of Contents for Accutorr 7
Page 1: ...Accutorr 7 Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 22: ...1 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 30: ...2 8 Accutorr 7 Operator s Manual Trend Screen 2 6 3 9 5 1 8 7 ...
Page 76: ...6 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 84: ...7 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 86: ...8 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 96: ...9 10 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 102: ...10 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 136: ...15 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 164: ...A 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 172: ...B 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 180: ...C 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 198: ...F 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 199: ......