Accutorr 7 Operator’s Manual
11 - 3
11.6
Measuring CO
2
CAUTION
•
Connect an exhaust tube to the gas outlet connector of the monitor to
remove the calibration gases to a scavenging system.
•
Connect the gas outlet to the scavenging system when performing the
measurement.
•
Check that the alarm limit settings are appropriate before taking
measurement.
To measure CO
2
, follow this procedure:
1.
Select an appropriate sampling line according to the patient category.
2.
Connect the sampling line to the CO
2
adapter that is installed on the monitor.
3.
Connect the other end of the sampling line to the patient.
◆
For intubated patients requiring an airway adapter, install the airway
adapter between the patient circuit and the ventilator Y-piece.
CO
2
adapter
Sampling line
Sampling line
Connect to the ventilator
Connect to the patient
Airway adapter
Summary of Contents for Accutorr 7
Page 1: ...Accutorr 7 Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 22: ...1 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 30: ...2 8 Accutorr 7 Operator s Manual Trend Screen 2 6 3 9 5 1 8 7 ...
Page 76: ...6 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 84: ...7 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 86: ...8 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 96: ...9 10 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 102: ...10 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 136: ...15 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 164: ...A 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 172: ...B 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 180: ...C 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 198: ...F 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 199: ......