D - 10
Accutorr 7 Operator’s Manual
D.2.5
Recorder Alarm Messages
Alarm Message
Clearable?
(Yes/No)
Alarm
Level
Cause
Solution
Recorder Init
Error
Yes
Low
An error occurred to
the module, or there
is a problem with
the communication
between the
module and the
monitor.
Restart the monitor.
If the problem still
exists, contact
Mindray or your
service personnel.
Recorder Selftest
Error
Yes
Low
Recorder
Unavailable
Yes
Low
Recorder Comm
Error
Yes
Low
Recorder Vlt High
No
Low
There is a problem
with the system
power supply.
Restart the monitor.
If the message is
prompted for
several times,
contact Mindray or
your service
personnel.
Recorder Vlt Low
No
Low
Recorder Head
Hot
No
Low
The recorder has
been working for
too long time.
Stop the recording
and resume the
recording until the
recorder’s print
head cools down.
Rec Head Wrong
Pos.
Yes
Low
The thermal head of
the recorder is in
wrong position.
Restore the control
lever of the recorder
to its previous
position.
Recorder out of
paper
Yes
Low
The recorder paper
is used up.
Replace with a new
paper roll.
Summary of Contents for Accutorr 7
Page 1: ...Accutorr 7 Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 22: ...1 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 30: ...2 8 Accutorr 7 Operator s Manual Trend Screen 2 6 3 9 5 1 8 7 ...
Page 76: ...6 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 84: ...7 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 86: ...8 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 96: ...9 10 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 102: ...10 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 136: ...15 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 164: ...A 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 172: ...B 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 180: ...C 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 198: ...F 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 199: ......