Accutorr 7 Operator’s Manual
6 - 11
6.10
Setting the Reminder Tone
When the alarm volume is set to zero, or alarm is reset, the monitor can issue a periodical
reminder tone. The interval of the reminder tone is 1 minute. You can switch on or off the
reminder tone.
1.
Select [
Main
]
→
[
Maintenance >>
]
→
[
User Settings >>
]
→
Enter required
password
→
Select [
Ok
] to access [
User Settings
] menu.
2.
Select [
Alarm Setup >>
] to access [
Alarm Setup
] menu.
3.
Set [
Reminder Tone
] to [
On
] or [
Off
].
6.11
Latching Alarms
The latching setting for physiological alarms defines how alarm indicators behave when
you do not acknowledge them.
■
If an alarm is latched, alarm indications remain presented even though alarm
conditions end, except that:
◆
The parameter reading and violated alarm limit stop flashing.
◆
The time when the alarm is last triggered is displayed after the alarm
message.
■
If an alarm is not latched, the alarm indications disappear as soon as the alarm
conditions end.
When the alarm system is reset, the latched physiological alarms are cleared.
To latch a physiological alarm,
1.
Select [
Main
]
→
[
Maintenance >>
]
→
[
User Settings >>
]
→
Enter required
password
→
Select [
Ok
] and then access [
User Settings
] menu.
2.
Select [
Alarm Setup >>
].
3.
Set [
Alarm Latch
] to [
High only
], [
High&Med
], [
All
] or [
Off
].
◆
[
High only
]: only high priority alarms are latched;
◆
[
Hi&Med
]: both high priority alarms and medium priority alarms are
latched;
◆
[
All
]: all alarms are latched; and
◆
[
Off
]: no alarm will be latched.
Only the unacknowledged physiological alarm can be latched. The latched alarms will
be cleared when the monitor enters the alarm reset state
.
WARNING
Summary of Contents for Accutorr 7
Page 1: ...Accutorr 7 Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 22: ...1 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 30: ...2 8 Accutorr 7 Operator s Manual Trend Screen 2 6 3 9 5 1 8 7 ...
Page 76: ...6 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 84: ...7 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 86: ...8 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 96: ...9 10 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 102: ...10 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 136: ...15 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 164: ...A 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 172: ...B 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 180: ...C 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 198: ...F 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 199: ......