Accutorr 7 Operator’s Manual
D - 7
D.2.3
Temperature Alarm Messages
Alarm Message
Clearable?
(Yes/No)
Alarm
Level
Cause
Solution
Temp Init Err
Yes
High
An error occurred to
the Temp module,
or there is a
problem with the
communications
between the
module and the
monitor, or Temp
calibration error.
Restart the monitor.
If the error remains,
contact Mindray or
your service
personnel.
Temp Comm Err
No
High
An error occurred to
the Temp module,
or there is a
problem with the
communications
between the
module and the
monitor.
Restart the monitor.
If the error remains,
contact Mindray or
service personnel.
Temp Alarm Limit
Err
No
High
The alarm limit of
Temp is changed
inadvertently.
Contact Mindray or
your service
personnel.
Warmup Timed
Out
Yes
Med
The initial probe
temperature in
measurement is too
high.
Cool the probe and
retry.
Warming Resistor
Err
No
Med
The thermal resistor
on the temperature
probe has an error
(Can not work
properly).
Replace
temperature probe.
Ambient Temp
Overrange
Yes
Med
The environmental
temperature is out
the range of the
monitor’s
measurement.
Change an
environment and
retry.
Temp Voltage Err
Yes
Med
The voltage is too
high or too low.
Check the power
supply.
Temp Prediction
Err
Yes
Low
The measuring
operation is
improper.
Retry the
measurement.
Summary of Contents for Accutorr 7
Page 1: ...Accutorr 7 Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 22: ...1 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 30: ...2 8 Accutorr 7 Operator s Manual Trend Screen 2 6 3 9 5 1 8 7 ...
Page 76: ...6 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 84: ...7 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 86: ...8 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 96: ...9 10 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 102: ...10 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 136: ...15 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 164: ...A 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 172: ...B 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 180: ...C 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 198: ...F 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 199: ......