Accutorr 7 Operator’s Manual
A - 9
A.7.3
NIBP
Standards
Meet standards of IEC80601-2-30, EN1060-1, EN1060-3, EN1060-4
and SP10
Technique
Oscillometry
Mode of
operation
Manual, Auto, STAT and Program
Auto mode
repetition
intervals
1, 2, 2.5, 3, 5, 10, 15, 20, 30, 60, 90 or 120 min
STAT mode cycle
time
5 min
Max
measurement
time
Adult, pediatric: 180 s
Neonate: 90 s
Heart rate range
3
0 to
30
0 bpm
Measurement
ranges (mmHg)
Adult
Pediatric
Neonate
Systolic
25 to 290
25 to 240
25 to 140
Diastolic
10 to 250
10 to 200
10 to 115
Mean
15 to 260
15 to 215
15 to 125
Accuracy
Max mean error:
5 mmHg
Max standard deviation: 8 mmHg
Static pressure
measurement
range
0 mmHg to 300 mmHg
Static pressure
measurement
accuracy
±3 mmHg
Resolution
1mmHg
Initial cuff
inflation
pressure range
(mmHg)
Adult:
Pediatric:
Neonate:
80 to 280
80 to 210
60 to 140
Summary of Contents for Accutorr 7
Page 1: ...Accutorr 7 Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 22: ...1 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 30: ...2 8 Accutorr 7 Operator s Manual Trend Screen 2 6 3 9 5 1 8 7 ...
Page 76: ...6 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 84: ...7 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 86: ...8 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 96: ...9 10 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 102: ...10 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 136: ...15 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 164: ...A 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 172: ...B 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 180: ...C 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 198: ...F 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 199: ......