Accutorr 7 Operator’s Manual
16 - 3
■
Isopropanol (70%)
■
Perform® classic concentrate OXY
16.1.3
Sterilization
Sterilization is not recommended for this equipment, related products, accessories, or
supplies unless otherwise indicated in the Instructions for Use that accompany the
products, accessories or supplies.
16.2
General Inspection
Before first use, follow these guidelines when inspecting the equipment:
■
Make sure that the environment and power supply meet the specifications
.
■
Inspect the equipment and its accessories for mechanical damage.
■
Inspect all power cords for damage, and make sure that their insulation is in good
condition.
■
Make sure that only specified accessories are applied.
■
Inspect if the alarm system functions correctly.
■
Make sure that the recorder functions correctly and the recorder paper meets the
requirements.
■
Make sure that the battery meets the performance requirements.
■
Make sure that the monitor is in good working condition
In case of any damage or abnormality, do not use the equipment. Contact the hospital’s
biomedical engineers or your service personnel immediately.
16.3
Maintenance and Testing Schedule
The following maintenance and tests, except for visual inspection, power on test,
touchscreen calibration, and battery check, should be carried out by the service
personnel only. Ensure the monitor is safety and performance tested by qualified service
personnel before initial installation, after repair or upgrade or during regularly
scheduled maintenance. Contact your service personnel if any maintenance is required.
Make sure to clean and disinfect (if required) the equipment before any test and
maintenance.
CAUTION
•
Care should be taken to change the settings in [User Settings >>] and
[Factory Maintenance >>] menus to avoid data loss
.
•
Service personnel should acquaint themselves with the test tools and make
sure that test tools and cables are applicable.
Summary of Contents for Accutorr 7
Page 1: ...Accutorr 7 Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 22: ...1 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 30: ...2 8 Accutorr 7 Operator s Manual Trend Screen 2 6 3 9 5 1 8 7 ...
Page 76: ...6 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 84: ...7 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 86: ...8 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 96: ...9 10 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 102: ...10 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 136: ...15 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 164: ...A 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 172: ...B 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 180: ...C 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 198: ...F 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 199: ......