Accutorr 7 Operator’s Manual
D - 9
XX represents a measurement or parameter label
.
CO2 SelfTest
Err
No
High
An error occured
during CO2
module
initialization.
Replace the
module.
CO2 Sensor
High Temp
No
Low
Ambient
temperature is
too high or there
is a module
failure.
Lower the
operating
temperature.
If the alarm
persists, the CO2
module may
have a failure.
Contact Mindray
or your service
personnel.
CO2 Zero
Failed
No
Low
The zeroing
failed.
Check the CO2
connections. If
the alarm
persists, contact
Mindray or your
service
personnel.
CO2 No
Filterline
Yes
Low
The sample line is
off or has been
disconnected.
Make sure that
the sample line is
connected.
CO2 FilterLine
Occluded
No
Low
The airway or
water trap is
occluded.
Check the airway
and remove the
occlusion.
XX Overrange
No
Low
The measured
parameter or
measurement XX
value is not
within the
specified range.
Contact Mindray
or your service
personnel.
Alarm Message
Clearable?
(Yes/No)
Alarm
Level
Cause
Solution
Summary of Contents for Accutorr 7
Page 1: ...Accutorr 7 Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 22: ...1 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 30: ...2 8 Accutorr 7 Operator s Manual Trend Screen 2 6 3 9 5 1 8 7 ...
Page 76: ...6 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 84: ...7 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 86: ...8 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 96: ...9 10 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 102: ...10 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 136: ...15 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 164: ...A 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 172: ...B 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 180: ...C 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 198: ...F 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 199: ......