D - 8
Accutorr 7 Operator’s Manual
D.2.4
CO
2
Alarm Messages
Temp SelfTest Err
No
High
An error occurs
during Temp
module
initialization.
Replace the module.
Temp Over High
Limit
No
High
The patient’s
temperature is too
high, or an error
occurs.
Reduce the patient’s
temperature, or
replace the module.
Temp Over Low
Limit
No
High
The patient’s
temperature is too
low, or an error
occurs.
Raise the patient’s
temperature, or
replace the module.
Temp No Probe
Yes
Med
The probe is
disconnected.
Reconnect the
probe.
Temp Probe
Misplaced
Yes
Med
The temperature
probe is not well
placed, or not
inserted into probe
well.
1
.
Check if the probe
well is installed.
2
.
Properly re-insert
the probe into
probe well.
Temp Measuring
Timeout
Yes
Med
The measuring time
is over 5 minutes in
Monitor mode.
Return the sensor to
the probe well, and
take a temperature
again.
Alarm Message
Clearable?
(Yes/No)
Alarm
Level
Cause
Solution
Alarm Message
Clearable?
(Yes/No)
Alarm
Level
Cause
Solution
CO2 Comm Err
No
High
An error occurred
to the CO2
module, or there
is a problem with
the
communication
between the
module and the
monitor.
Restart the
monitor. If the
error remains,
contact Mindray
or your service
personnel.
CO2 Comm
Abnormal
No
High
CO2 Comm
Stop
No
High
CO2 Init Err
No
High
Summary of Contents for Accutorr 7
Page 1: ...Accutorr 7 Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 22: ...1 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 30: ...2 8 Accutorr 7 Operator s Manual Trend Screen 2 6 3 9 5 1 8 7 ...
Page 76: ...6 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 84: ...7 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 86: ...8 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 96: ...9 10 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 102: ...10 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 136: ...15 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 164: ...A 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 172: ...B 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 180: ...C 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 198: ...F 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 199: ......