Accutorr 7 Operator’s Manual
A.7.2
PR
PR from Masimo SpO
2
Module
PR from Nellcor SpO
2
Module
PR from NIBP Module
Module
PR High Limit
PR Low Limit
Step
(bpm)
Masimo SpO
2
Module
(low limit +2) to 240
25 to (high limit-2)
1
Nellcor SpO
2
Module
(low limit +2) to 300
20 to (high limit-2)
NIBP Module
(low limit +2) to 300
30 to (high limit-2)
Measurement range
25 to 240 bpm
Resolution
1 bpm
Response time
≤ 20 s (PR value sudden change within 25 – 240 bpm)
Accuracy
±3 bpm (without motion)
±5 bpm (with motion)
Refreshing rate
1 s
Measurement range
20 to 300 bpm
Resolution
1 bpm
Response time
≤ 30 s (PI > 0.3, no disturbance, PR value sudden change
within 25 – 250 bpm)
Accuracy
20 to 250 bpm: ±3 bpm
251 to 300 bpm, not specified
Refreshing rate
1 s
Measurement range
30 to 300 bpm
Resolution
1 bpm
Accuracy
±3 bpm or ±3%, whichever is greater
Summary of Contents for Accutorr 7
Page 1: ...Accutorr 7 Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 22: ...1 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 30: ...2 8 Accutorr 7 Operator s Manual Trend Screen 2 6 3 9 5 1 8 7 ...
Page 76: ...6 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 84: ...7 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 86: ...8 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 96: ...9 10 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 102: ...10 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 136: ...15 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 164: ...A 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 172: ...B 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 180: ...C 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 198: ...F 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 199: ......