6 - 6
Accutorr 7 Operator’s Manual
WARNING
•
When the alarm sound is switched off, the equipment will give no audible
alarm tones even if a new alarm occurs. Therefore the user should be very
careful about whether to switch off the alarm sound or not.
•
Do not rely exclusively on the audible alarm system for patient monitoring.
Adjustment of alarm volume to a low level may result in a hazard to the
patient. Always keep the patient under close surveillance.
6.5.4
Adjusting Alarm Limits Automatically
The monitor can automatically adjust alarm limits according to the measured vital signs,
using the auto limits function. When auto limits are selected, the monitor calculates safe
auto limits based on the latest measured values.
To get accurate auto alarm limits, you need to collect a set of measured vital signs as a
baseline. Then, in the main menu, select [
Main
]
→
[
Alarm Setup >>
]
→
[
Auto Set
]
→
Select [
Ok
] in the pop-up window. The monitor will create new alarm limits based on the
measured values.
Before applying these automatically created alarm limits, confirm if they are appropriate
for your patient. If not, you can adjust them manually.
The monitor calculates the auto limits based on the following rules.
Parameter
Low alarm Limit
High alarm Limit
Auto-set alarm
limit Range
Adult/
Pediatric
Neonate
Adult/
Pediatric
Neonate
SpO
2
Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
measurement
range
PR
(PR×0.8) or 40
bpm,
whichever is
greater
(PR-30) or 90
bpm,
whichever is
greater
(PR×1.25) or
240 bpm,
whichever is
smaller
(PR+40) or
200 bpm,
whichever is
smaller
Adult/Pediatric:
35 to 240 bpm
Neonate:
55 to 225 bpm
Summary of Contents for Accutorr 7
Page 1: ...Accutorr 7 Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 22: ...1 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 30: ...2 8 Accutorr 7 Operator s Manual Trend Screen 2 6 3 9 5 1 8 7 ...
Page 76: ...6 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 84: ...7 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 86: ...8 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 96: ...9 10 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 102: ...10 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 136: ...15 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 164: ...A 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 172: ...B 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 180: ...C 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 198: ...F 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 199: ......