Accutorr 7 Operator’s Manual
D - 1
D
Alarm Messages
This chapter lists only the most important physiological and technical alarm messages.
Some messages appearing on your monitor may not be included. In the “Solution”
column, corresponding solutions are given instructing you to troubleshoot problems. If
the problem persists, contact your service personnel.
■
“*” means the alarm level is user-adjustable.
■
XX represents a measurement or parameter label, such as NIBP, PR, etc.
D.1
Physiological Alarm Messages
Alarm
Message
Alarm Priority
Cause
Solution
XX Too High
Med*
XX value exceeds the
upper alarm limit.
Check the patient’s
condition and check if the
patient category and
alarm limit settings are
correct.
XX Too Low
Med*
XX value is lower than the
lower alarm limit.
SpO
2
Desat
High
The SpO
2
value has fallen
below the desaturation
alarm limit.
Check the patient’s
condition and check if the
alarm limit settings are
correct.
No Pulse
High
The pulse signal was so
weak that the monitor
cannot perform pulse
analysis.
Check the patient’s
condition, SpO
2
sensor
and measurement site.
Check
breathing. No
CO
2
Changes
Detected
High
The patient stops
breathing, or the
respiration signal was so
weak that the monitor
cannot perform
respiration analysis.
Check the patient’s
condition, module and
patient connections.
Summary of Contents for Accutorr 7
Page 1: ...Accutorr 7 Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 22: ...1 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 30: ...2 8 Accutorr 7 Operator s Manual Trend Screen 2 6 3 9 5 1 8 7 ...
Page 76: ...6 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 84: ...7 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 86: ...8 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 96: ...9 10 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 102: ...10 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 136: ...15 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 164: ...A 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 172: ...B 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 180: ...C 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 198: ...F 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Page 199: ......