Additional Imaging Modes
5
Keysight 5500 SPM User’s Guide
5-10
10 nA/V (red), 1 nA/V (blue) or 0.1 nA/V (green). See
Figure 7-10
CSAFM nose assembly
Platinum-coated, conductive tips are required for CSAFM imaging.
Because an electrode must be attached to the sample, a sample plate is
also required.
To image in CSAFM Mode:
1
Begin with the steps you learned in
a
Insert the nose assembly into the scanner.
b
Load a probe into the nose assembly.
c
Place the scanner in the microscope base and connect its cables.
d
Align the laser on the cantilever.
e
Insert and align the detector.
2
Prepare the sample and place it on a sample plate. The sample must
be electrically isolated from the sample plate. The particular
mounting arrangement will depend on the sample type and size.
3
Attach an electrode from the sample plate to the sample. A length of
copper wire works well as the electrode. Lift the spring-loaded
electrode clip on the sample plate and insert the electrode under it
(
on page 7). Connect the electrode to the sample. Check
CAUTION
The CSAFM nose assembly cannot be used for imaging in liquid.