MAC III
5
Keysight 5500 SPM User’s Guide
5-37
procedure is for AM mode operation. Note the tip should
not
be engaged
at this point.
1
Choose
Mode > KFM > AM
, then choose
Controls > AC Mode
to
open the KFM Controls window.
2
Engage the tip and optimize the image.
3
Set parameters in the KFM tab as shown in
for initial
settings.
Figure 11-18
KFM tab initial settings (left) for AM mode operation.
Settings on the right are after the Drive (%) check box has been
selected and the Optimize button has been clicked.
4
Select the
Drive (%)
check box.
This will turn on the AC bias to the tip. If the Amplitude 2 parameter (in
the lower portion of the KFM tab) is less than 0.100 V, increase the
Drive (%) or Gain 2.