Keysight 5500 SPM Components
5
Keysight 5500 SPM User’s Guide
5-15
AFM Controller
) provides the high voltage to the
piezoes and other control functions. Model N9605A is standard; Model
N9610A provides optional closed-loop scanning control and extra data
channels required for modes such as EFM and KFM.
Figure 2-13
AFM Controller (Model N9610A)
The
Aux In
BNC input is added on to the current signal, which is used
in STM and CSAFM. The
External Input
BNC to the Z piezo is
limited to ±10 V and the input is scaled up by the controller to ±215 V.
The
External Sync
BNC is a signal that is sent each time a frame, line,
or pixel is completed. This can be used in Custom Spectroscopy or