Electrochemistry
5
Keysight 5500 SPM User’s Guide
5-15
3
Guide the Pt wire through the small hole in the nose cone.
Figure 16-12
Assembled SECM nose cone with wire threaded through
hole in nose cone.
NOTE
The wire is insulated with a thin Teflon coating. Be careful not to
damage the coating as a large leakage current could result, leading to an
inability to run SECM experiments.
4
Position the end of the wire between the base contact and set screw
in the nose cone as seen in
.
Figure 16-13
Arrow indicates point of contact for probe wire.