MAC III
5
Keysight 5500 SPM User’s Guide
5-8
Lock-in 1
Drive Output
This is the drive output of Lock-in Amplifier 1 and can be used as the AC stimulus of
an external experiment. The output is nominally in the range of ±10 V. A DC bias can
also be added by entering a voltage for the Drive Offset in the Lock-in 1 tab in the
Advanced AC Mode Controls window.
Lock-in 1
Input
This is the auxiliary input of Lock-in Amplifier 1. The nominal input range is 2.5 V
peak to peak AC within the range of ±10 V DC. To use this input to Lock-in 1, open
the Advanced AC Mode Controls window, select the Lock-in 1 tab, then select
Aux
from the Input dropdown menu.
This can also be used as an external input to the controller. While this is in use it
disables the normal outputs of the Lock-in. To use this as an external input to the
controller, select the Lock-in 1 tab in the Advanced AC Mode Controls window and
select the
Y Component From Aux
check box to enable the Auxiliary signal to
replace the Y component. Then in the Output tab, select
Y Component 1
as an output
to one on the controller auxiliary inputs.
Lock-in 2
Drive Output
This is the drive output of Lock-in Amplifier 2 and can be used as the AC stimulus of
an external experiment. The output is nominally in the range of ±10 V. A DC bias can
also be added by entering a voltage for the Drive Offset in the Lock-in 2 tab in the
Advanced AC Mode Controls window.
Lock-in 2
Input
This is the auxiliary input of Lock-in Amplifier 2. The nominal input range is 2.5 V
peak to peak AC within the range of ±10 V DC. To use this input to Lock-in 2, open
the Advanced AC Mode Controls window, select the Lock-in 2 tab, then select
Aux
from the Input dropdown menu.
This can also be used as an external input to the controller. While this is in use it
disables the normal outputs of the Lock-in. To use this as an external input to the
controller, select the Lock-in 2 tab in the Advanced AC Mode Controls window and
select the
Y Component From Aux
check box to enable the Auxiliary signal to
replace the Y component. Then in the Output tab, select
Y Component 2
as an output
to one on the controller auxiliary inputs.
Lock-in 3
Drive Output
This is the drive output of Lock-in Amplifier 3 and can be used as the AC stimulus of
an external experiment. The output is nominally in the range of ±10 V. A DC bias can
also be added by entering a voltage for the Drive Offset in the Lock-in 3 tab in the
Advanced AC Mode Controls window.