Additional Imaging Modes
5
Keysight 5500 SPM User’s Guide
5-4
sensitivity. The optional modules are 1 nA/V and 0.1 nA/V. Other
values are available upon request.
Figure 7-4
STM scanner preamp cable and module
To exchange these modules, please follow the procedure described
below.
CAUTION
Do not pull the cable while unplugging the module. Hold the 4-pin con-
nector at the end of the cable firmly, then unplug the module as shown in
1
Remove the four screws on the tip holder piece.
2
Pull out the preamp module carefully.
3
Hold on to the 4-pin connector and unplug the module from the
cable.
4
Replace the module and plug it back onto the 4-pin connector.
5
Tighten the module back onto the scanner.