Setting Up the Keysight 5500 SPM
5
Keysight 5500 SPM User’s Guide
5-6
specified noise level, the vibration isolation chamber installation surface
must meet Criterion D.
Figure 3-2
Vibration criterion curves and ISO guidelines
Temperature and Humidity Variation
Changes in temperature and humidity will affect both resolution and
repeatability of imaging. Temperature variation should be limited to ±2
degrees Fahrenheit. Humidity variation should not exceed ±20 % RH.
Locating the instrument away from vents and air handlers will help meet
this goal.