Scanner Maintenance and Calibration
5
Keysight 5500 SPM User’s Guide
5-7
over time. A large amount of decrease takes place during the first few
hours of use. Therefore, scanners are burned in before initial calibration.
After this initial burn-in, the aging rate is very slow; however,
calibration once or twice per year is still recommended.
Creep
At a constant applied voltage the piezo position will change slightly
over time, most noticeably at the beginning of a scan or after a change
on scan location. Creep appears as elongation of the feature in the
direction of the offset because of this slow movement (
Figure 8-3
Scanner creep