AC Modes
5
Keysight 5500 SPM User’s Guide
5-5
Figure 6-3
ACAFM tab in AC Tune window.
7
Set the Drive (%) to
10 %
. This is the initial amplitude of the AC
drive signal, stated as a percentage (0-100 %) of the maximum
available 10 V.
8
In the Auto Tune tab (
) of the AC Tune
window, enter the
Start
and
End
frequencies (in kHz) for the tuning sweep. For a new
or unknown cantilever, use the stated minimum and maximum
frequencies given on the storage box. If the resonance frequency
range is known more exactly, a smaller range can be used to speed
the tuning process.
NOTE
If the system is using the AC source in the HEB, the Setpoint needs to
be set to zero to perform Auto Tune. If the system has an external
AC/MAC Mode Module, the Setpoint can be any value during Auto
Tune.