Additional Imaging Modes
5
Keysight 5500 SPM User’s Guide
5-7
). Connect the electrode to the sample,
ensuring good contact.
Figure 7-7
Sample on plate with electrode attached
6
Place the sample plate on the microscope.
7
Plug the 3-pin EC connector of the EC/MAC cable into the 3-pin
socket on the sample plate. Plug the other end of the cable into the
EC/MAC socket on the microscope.
8
In PicoView, choose
Mode > STM
.
9
In the Servo
window enter the Bias Voltage (
). Typical
values are
50-200 millivolts
(0.05-0.200 V). A positive bias
indicates current flow
from
the tip
to
the sample, and vice versa for
negative bias.
NOTE
The sample plate cable can transfer low levels of vibration to the
sample. During very high resolution imaging this can affect images
quality. We recommend first plugging the sample plate cable to the
flexible 3-wire umbilical included with the sample plate. The umbilical
should then be plugged in to the microscope base. The umbilical’s
individual wires tend to reduce the transfer of vibration.