AC Modes
5
Keysight 5500 SPM User’s Guide
5-13
the tip. The standard sample plate with the magnet in the center for
holding sample pucks cannot be used for Top MAC imaging due to the
interference from the magnet in the sample plate.
Figure 6-9
Top MAC nose assembly
Q Control
An oscillating cantilever in AC Mode is influenced by complex
interaction forces between it and the surface. By carefully setting system
parameters the system can be made to operate entirely in the regime of
net-attractive forces, thereby reducing the effect of the probe tip on the
sample. The range in which the parameters have to be adjusted can be
narrow, however, making it difficult to maintain in real operation.
Q Control is a method that reduces damping of the system, increasing
the quality factor of the oscillating cantilever. Enhanced resonance
allows imaging with very low force and high phase sensitivity. The