Closed-Loop Scanners
5
Keysight 5500 SPM User’s Guide
5-15
Once the plot looks like the example above in
Offset and Gain are properly set and the Z Sensor is usable over its
entire range.
17
Select the
CL Sweep
check box. The plot should change from the
one above to that shown in
Figure 9-8
Spectroscopy sweep with closed loop On
18
In the Scanner Setup Closed Loop tab, adjust the Z Range setting to
maximize the range. A typical starting value is 19 V.
NOTE
The Z piezo is highly sensitive to changes in Offset and Gain. To change
these values in smaller increments, click the mouse in the Offset or Gain
box. Hold down the
Ctrl
key while pressing the
Up
or
Down
arrows to
change the parameter in 0.01 steps. Hold down the
Ctrl
and
Shift
keys
while pressing the
Up
or
Down
arrow keys to change the values in
0.001 increments.