Additional Imaging Modes
5
Keysight 5500 SPM User’s Guide
5-26
6
Choose
Controls > Advanced > AC Mode.
The EFM Controls
window will open (
).
Figure 7-14
EFM Controls window
7
In the Main tab, set up the Lock-in 1 AC signal which drives the
cantilever oscillation:
a
Set the Drive % to 10 %.
b
Set the Gain to
x1
(the amplitude times 1).
8
Choose
Controls > AC Mode Tune
to open the AC Tune
window.
9
Tune the drive signal to the resonance frequency of the cantilever:
a
In the
AC Tune
window, enter the
Start
and
End
frequencies (in
kHz) for the tuning sweep, typically 20-120 kHz.
b
Set the Peak Amplitude to
2.5
volts.
c
Enter an
Off Peak
value to offset the oscillation frequency from
the cantilever’s resonance frequency. A typical value is
-0.100 kHz
.
d
Click the
Auto Tune
button. The system will sweep through the
range of frequencies, locating the peak oscillation amplitude
within the range. The AC signal oscillation will be set to this
value plus the specified Offset.
NOTE
Q Control can be used with EFM Mode to improve the resonance peak
and image quality. See