Scanner Maintenance and Calibration
5
Keysight 5500 SPM User’s Guide
5-10
6
Choose
Scanner > Edit
to open the Scanner Setup dialog box. Select
the Open Loop tab.
Figure 8-5
Scanner Setup dialog box
X Non-Linearity
In the Realtime Images window, choose
Tools > Horizontal Cross
Section
. Use markers to report the dimensions between sets of features
at either end of the scan range (
Figure 8-6
Horizontal cross-section tool used to check non-linearity.
Two sets of cursors are shown.