Additional Imaging Modes
5
Keysight 5500 SPM User’s Guide
5-6
Final Sensitivity Calculation
To calculate the final sensitivity, use this equation:
For example, if the Gain switch is set to 10 and the module has a
10 nA/V preamp, the final sensitivity will be 1 nA/V.
STM Imaging Procedure
The procedure for STM imaging is as follows:
1
If you are using the multi-purpose scanner, insert the nose assembly
into the scanner (see
2
Insert a tip into the nose assembly or scanner. Grasp the tip with a
tweezers, then insert it into the hollow tube until it protrudes
approximately 2 mm (
Figure 7-6
Inserting STM tip wire in scanner
3
Place the scanner in the scanner bracket on the microscope.
4
Prepare the sample and place it on a sample plate. The sample must
be electrically isolated from the sample plate. The particular
mounting arrangement will depend on the sample type and size.
5
Attach an electrode from the sample plate to the sample. Lift the
spring-loaded electrode clip on the sample plate and insert the
Final Sensitivity
PreampSensitivity
Gain
=