Zeiss LSM 710

 

 

1  LSM 710 – Quick Guide                                                MFPL BioOptics - Light Microscopy 

 

Startup 

1.

 

Turn on the Main Switch (see image below, #1). 

2.

 

Turn on the Systems/PC and Components Switches (see image below, #2). 

3.

 

Power up the X-Cite short arc lamp (see image below, #3). 
Note: the lamp should not be turned on/off in quick succession. Always wait ca. 20 min. 

4.

 

Turn the key of the Argon laser power supply 90° (clockwise) from position 0  to 1 (see below, #4). 

 

5.

 

Turn on the PC (no image shown). 

6.

 

After about 5 min, switch the Lasos Argon laser control unit from 'idle' to 'run' (see image below, 
#6). Wait until the green and/or red LED on the Lasos control unit lights up. If the red LED shows 
up, turn the control knob counterclockwise until only the green LED is on. 
Note: if you use only 488 nm at low power, you may leave it in 'idle' mode all the time. 

 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

7.

 

Start the software: double-click on the ZEN icon. In the Login dialog, select 'Start System'. Once 
ZEN is loaded, select the 'View' menu and check 'Show all (global)'. 
 