Operation
Plug in the power cord. The
green
signal lamp indicates that the unit is switched on. Until the in-
put pressure from the connected cylinders, A and B, drops below
0,6 bar
, the
red
signal lamp re-
mains on. The yellow signal lamp indicates the cylinder currently selected.
Open the valve on the gas cylinders and set the pressure regulator on the cylinders for an initial
pressure between
0,9 to 1,1 bar
. The
red
signal lamps should go out. Press the
blue
buttons to
select the desired cylinder (either A or B), and to switch it to the output. After a cylinder has been
manually selected, selection of the remaining cylinder is prevented for approx. 5 seconds. When
the connected device (i.e. a gassed incubator) begins draining gas from the selected cylinder, a
pressure drop may occur at that cylinder's pressure reducer. For this reason, we recommend ad-
justing the output pressure at the pressure reducers to
0,9 to 1,1 bar
when gas is being removed.
An increase in input pressure once gas has been removed has no negative effect on the operati-
on of the unit as long as the maximum pressure of
2 bar
is not exceeded.
If the pressure in the selected cylinder drops below
0,6 bar
for more than 1 minute, the gas cylin-
der monitor automatically switches the supply to the second gas cylinder, and the
red
signal lamp
of the empty cylinder comes on. Simultaneously, the acoustic alarm is switched on and the
alarm output becomes active. Press the "Acknowledge Fault" button to turn the acoustic alarm
off. Should the pressure in the empty cylinder again rise above
0,7 bar
after the unit has already
switched over to the new cylinder, the unit will not automatically switch back to the previously
empty cylinder. However, the empty cylinders
red
signal lamp will go out and its
yellow
signal
lamp will begin flashing, so that the empty cylinder can be identified. After the empty cylinder has
been replaced, the
red
or flashing
yellow
signal lamps will turn off and the alarm output will no
longer be activated.
If, one the alarm has been acknowledged, the empty cylinder is not replaced with a full one, the
acoustic alarm is again activated once the second, full cylinder becomes empty. Whenever the
pressure in one or both cylinders drops below 0.6 bar for more than 1 minute, the alarm output
will be continuously activated.
Possible error messages
1.
Green signal lamp flashing
Switching between gas cylinders is performed by a built-in magnetic valve (bi-stable ma-
gnetic valve). This prevents a switchover from occurring if there is a power failure. The
unit's electronics determines and monitors the position of the magnetic valve. If the unit
does
not
detect any response from the magnetic valve after switching over, the procedure
is repeated up to 10 times. If the magnetic valve fails to switch over despite repeated at-
tempts to do so, the green signal lamp will flash. A flashing
green
lamp can only be reset
by unplugging the unit from the mains power supply.
2.
A yellow signal lamp is flashing
Pressure dropped below
0,6 bar
for longer than one minute. However, after switching over
to a full cylinder, the pressure climbed above
0,7 bar
. This condition can occur if the pres-
sure at the cylinder ouptut drops below
0,6 bar
as gas is being removed. In this case, ad-
just the pressure at the pressure reducer so that it lies between
0.9 to 1.1 bar
when gas is
drained from the cylinder.
3.
Both yellow signal lamps are flashing
This is the same problem as described under 2. above. Here however, the pressure in the
second cylinder has also dropped below 0.6 bar for more than one minute after the switch-
over. In this case, adjust the pressure at both pressure reducers so that it lies between
0,9
to 1,1 bar
when gas is drained from the cylinder.
5.
SPECIFICATIONS
50 047 292
7/9
Summary of Contents for HERAcell 150
Page 82: ...CO2 Incubator HERAcell 150 Operating Instructions 82 16 Certificate of decontamination...
Page 83: ...CO2 Incubator HERAcell 150 Operating Instructions 83 16 Certificate of decontamination...
Page 84: ...Internet http www kendro com...
Page 85: ......
Page 87: ...Valid 10 2005 50047292 B Gas Cylinder Change Over Unit GM 2 Operating instructions GB...
Page 106: ......
Page 152: ......
Page 153: ......
Page 154: ......
Page 155: ......
Page 156: ......
Page 157: ......
Page 158: ......
Page 159: ......
Page 160: ......
Page 161: ......
Page 162: ......
Page 163: ......
Page 164: ......
Page 165: ......
Page 166: ......
Page 167: ......
Page 168: ......
Page 169: ......
Page 170: ......
Page 171: ......
Page 172: ......
Page 173: ......
Page 174: ......
Page 175: ......
Page 220: ......
Page 221: ......
Page 222: ......
Page 223: ......
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 237: ...Version 04 2004 50078754 Accessories for CO2 incubators Illustrations and describtion...
Page 249: ...Footprint HERAcell...
Page 250: ...Footprint HERAcell...