Incubator HERAcell
®
150 Operating Instructions
47
6.
Handling and control
6.10
Activating the auto-start routine
The auto-start function is an automated routine for the start and the subsequent
adjustment of the CO
2
measuring system. After the start, the device control adjusts
the temperature to the set nominal value while humidity is generated. When tem-
perature and relative humidity have reached constant values, the CO
2
measuring
system is automatically adjusted to these values, and the work space is supp-
lied with the preset quantity of CO
2
.
NOTE – Application of the routine
To ensure that the specified accuracy of the
CO
2
measuring system is maintained, the de-
vice should always be started using the auto-
start routine after the nominal temperature
setting has been changed by more than 1° C
or after extended interruptions of the operation
of the device. The auto-start routine should be
run at least every three months on the occa-
sion of cleaning and maintenance works.
Running the routine usually takes 5 to 7 hours.
At low room temperatures and when the device
is cold, it may take up to 10 hours until the
auto-start routine has been completed. If the
glass door is opened or if the power supply of
the device is interrupted while the routine is
running, the routine is interrupted and rerun
after the glass door has been closed and after
the power supply has been reestablished.
At the start of the auto-start routine, the work
space atmosphere must consist only of ambi-
ent air. The floorpan must be filled with a suf-
ficient quantity of water!
1. Open both doors until the acoustic alarm sounds after 30 seconds:
All current
actual values
flash at the displays, the ”door” LED illumi-
nates, after 30 seconds the acoustic alarm sounds.
2. Enter nominal values:
See sections 6.4 / 6.8.
3. Activate the auto-start routine:
Keep the
key depressed for 5 seconds.
The ”auto-start” LED flashes.
Summary of Contents for HERAcell 150
Page 82: ...CO2 Incubator HERAcell 150 Operating Instructions 82 16 Certificate of decontamination...
Page 83: ...CO2 Incubator HERAcell 150 Operating Instructions 83 16 Certificate of decontamination...
Page 84: ...Internet http www kendro com...
Page 85: ......
Page 87: ...Valid 10 2005 50047292 B Gas Cylinder Change Over Unit GM 2 Operating instructions GB...
Page 106: ......
Page 152: ......
Page 153: ......
Page 154: ......
Page 155: ......
Page 156: ......
Page 157: ......
Page 158: ......
Page 159: ......
Page 160: ......
Page 161: ......
Page 162: ......
Page 163: ......
Page 164: ......
Page 165: ......
Page 166: ......
Page 167: ......
Page 168: ......
Page 169: ......
Page 170: ......
Page 171: ......
Page 172: ......
Page 173: ......
Page 174: ......
Page 175: ......
Page 220: ......
Page 221: ......
Page 222: ......
Page 223: ......
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 237: ...Version 04 2004 50078754 Accessories for CO2 incubators Illustrations and describtion...
Page 249: ...Footprint HERAcell...
Page 250: ...Footprint HERAcell...