CO
2
Incubator HERAcell
®
150 Operating Instructions
53
6.
Handling and control
6.13
Gas guard failure display
When the pressure in one of the two gas cylinders drops below the limit value of
0.6 bar for more than two minutes, the gas guard switches the gas supply over to
the other cylinder (filling pressure > 0.6 bar).
The response time of two minutes prevents the gas minitor from switching over
upon momentary pressure fluctuations (e.g. when opening device doors).
The failure appears alternatingly as „actual value“ and „error message“ (see be-
low) on the display of the detected process gas.
If the filling pressure is below the limit value in both cylinders, the audible alarm
is triggered, the potential-free contact is activated, and a visual alarm (flashing
failure code) appears on the display of the detected process gas.
Error messages (Example: CO
2
supply):
Pressure drop in gas cylinder A (pressure < 0.6 bar)
and
Pressure drop in gas cylinder B (pressure < 0.6 bar)
and
Pressure drop in both monitored gas cylinders A and B (pressure < 0.6 bar)
and
Code
Description
Cause
Repair
300
O
2
below nominal value
Act. val. < nom. val. –1 %
• No O
2
• Prepressure low
• Supply line blocked
Check gas supply:
• Connect new gas cylinder
• Raise prepressure to 1 bar max.
• Check supply line to device
301
O
2
above nominal value
Act. Val. > nom. Val. +1 %
• Prepressure high
Check gas supply:
• Reduce prepressure to 1 bar
304
O
2
sensor faulty
Sensor circuit open/shorted Contact Service
306
No O
2
supply
Both O
2
cylinders empty
Replace one or both O
2
cylinders
307
O
2
gas monitor failure
O
2
gas monitor faulty or not
connected
Check connection of O
2
gas monitor at
cylinders, contact Service as required
400
Water level low
Water level ³ 1 min below
lower limit
Check water level and refill water as
required
500
Temperature ContraCon
routine below nom. val.
Actual value < 85 °C
Repeat decontamination routine; contact
Service, as required
501
Temperature ContraCon
routine above nom. val.
Actual value > 95 °C
Contact Service
502
Failure in ContraCon
routine
Power failure during heating
or holding phase
Silence alarm by pressing key "90 °C“ (2
times 5 sec), then restart routine
Summary of Contents for HERAcell 150
Page 82: ...CO2 Incubator HERAcell 150 Operating Instructions 82 16 Certificate of decontamination...
Page 83: ...CO2 Incubator HERAcell 150 Operating Instructions 83 16 Certificate of decontamination...
Page 84: ...Internet http www kendro com...
Page 85: ......
Page 87: ...Valid 10 2005 50047292 B Gas Cylinder Change Over Unit GM 2 Operating instructions GB...
Page 106: ......
Page 152: ......
Page 153: ......
Page 154: ......
Page 155: ......
Page 156: ......
Page 157: ......
Page 158: ......
Page 159: ......
Page 160: ......
Page 161: ......
Page 162: ......
Page 163: ......
Page 164: ......
Page 165: ......
Page 166: ......
Page 167: ......
Page 168: ......
Page 169: ......
Page 170: ......
Page 171: ......
Page 172: ......
Page 173: ......
Page 174: ......
Page 175: ......
Page 220: ......
Page 221: ......
Page 222: ......
Page 223: ......
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 237: ...Version 04 2004 50078754 Accessories for CO2 incubators Illustrations and describtion...
Page 249: ...Footprint HERAcell...
Page 250: ...Footprint HERAcell...