Service Handbuch HERAcell 150
Service Manual HERAcell 150
C
202613
D.Dornseiff
06/26/03
Dokumentnr./Document No.
Seite/Page
Datei/File
50051109_04_EN_C.doc
50051109 / C
8/ 9
Weitergabe sowie Vervielfältigung dieser Unterlage, Verwertung oder Mitteilung ihres Inhalts nicht gestattet, soweit nicht ausdrücklich zugestanden. Zuwiderhandlungen ver-
pflichten zu Schadensersatz. Alle Rechte für den Fall der Patenterteilung oder Gebrauchsmustereintragung vorbehalten. © Kendro Laboratory Products GmbH, Germany
Sensor-specific data storage, NVRAM.
All sensor-block-specific data are stored in the NVRAM on the sensor
board. The master processor reads/writes this information via the PC bus.
Main board P003
(also main board or main PCB)
Provides the following functions:
Mains power connection:
Device power socket with integrated fuse holder
Power supply creation:
The combinational circuit component (input voltage range ~ 80 - 270 V)
generates the 12 V DC low power voltage for the electronic components.
Voltage is displayed by an LED. Power for the auto-zero air pump is derived
from the main PCB’s power supply.
Signal processing, water level sensor
Temperature and CO
2
calculation and regulation, O
2
regulation:
The measured values from the sensor board are adjusted by the offset and am-
plification, and corresponding set values are generated. Control of all software
routines. The O
2
concentration is calculated by a separate O
2
board, described in
more detail below. O
2
regulation is also a function of the main board.
Actuator controls:
All heating system actuators are located on the main PCB.
CO
2
solenoid activation:
The 12 VDC CO
2
solenoid is connected directly to the board.
Sample protection:
In case of overtemperature, a special software routine linked to a universally
switching relay assumes the temperature regulating function. This upper tem-
perature limit band (set value +1
C) acts as the trigger band for this controller.
The software attempts to readjust the unit to the specified desired temperature
value.
Temperature limiting during the decontamination routine:
If the upper temperature limit band (set value + 5
C) is exceeded during the
decontamination routine, a universal relay shuts the unit's heating system
down permanently.
Summary of Contents for HERAcell 150
Page 82: ...CO2 Incubator HERAcell 150 Operating Instructions 82 16 Certificate of decontamination...
Page 83: ...CO2 Incubator HERAcell 150 Operating Instructions 83 16 Certificate of decontamination...
Page 84: ...Internet http www kendro com...
Page 85: ......
Page 87: ...Valid 10 2005 50047292 B Gas Cylinder Change Over Unit GM 2 Operating instructions GB...
Page 106: ......
Page 152: ......
Page 153: ......
Page 154: ......
Page 155: ......
Page 156: ......
Page 157: ......
Page 158: ......
Page 159: ......
Page 160: ......
Page 161: ......
Page 162: ......
Page 163: ......
Page 164: ......
Page 165: ......
Page 166: ......
Page 167: ......
Page 168: ......
Page 169: ......
Page 170: ......
Page 171: ......
Page 172: ......
Page 173: ......
Page 174: ......
Page 175: ......
Page 220: ......
Page 221: ......
Page 222: ......
Page 223: ......
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 237: ...Version 04 2004 50078754 Accessories for CO2 incubators Illustrations and describtion...
Page 249: ...Footprint HERAcell...
Page 250: ...Footprint HERAcell...