Incubator HERAcell
®
150 Operating Instructions
19
Fig. 3
Stacking
Fig. 4
Tray installation
3.
Installation of the device
3.5
Stacking
Fig. 3:
Two HERAcell
®
150 devices can be stacked on top of
each other. The upper device is secured to the lower device
by inserting the device stands [1] into the stacking elements
[2] on the device ceiling.
If the devices are placed onto mobile racks, ensure that the
rollers [3] are secured by an arresting device during operati-
on. For reasons of stability, the rollers should be oriented to
the front.
NOTE – Transporting stacked devices
The stacking elements are anot connecting
elements. Therefore, the transport of stacked
devices on sloped surfaces is not allowed.
3.6
Retrofitting/Modifications
The outer door and the glass door can be equipped with left
or right side fasteners. The door fastening can also be reversed
later.
The standard version can also be retrofitted with a three-ele-
ment split gas bezel by replacing the glass door with the gas
bezel.
NOTE – Modifications
Retrofittings and modifications must only be
performed by the Technical Service of Kendro
Laboratory Products.
3.7
Tray
Fig. 4: After the device has been installed, the tray is inserted
between the two guide straps below the bottom of the device.
1. Open the outer door of the device all the way to the stop.
2. Fold the guide straps [1] all the way down.
3. Position tray [4] with vertical edge [2] below cable guide [3].
The quick reference [5] faces the front of the device.
4. Lift tray slightly and tilt it downward.
5. Push tray in until resistance is observed, then push tray in
all the way to the stop by pushing onto guide straps.
4
3
2
1
5
4
Summary of Contents for HERAcell 150
Page 82: ...CO2 Incubator HERAcell 150 Operating Instructions 82 16 Certificate of decontamination...
Page 83: ...CO2 Incubator HERAcell 150 Operating Instructions 83 16 Certificate of decontamination...
Page 84: ...Internet http www kendro com...
Page 85: ......
Page 87: ...Valid 10 2005 50047292 B Gas Cylinder Change Over Unit GM 2 Operating instructions GB...
Page 106: ......
Page 152: ......
Page 153: ......
Page 154: ......
Page 155: ......
Page 156: ......
Page 157: ......
Page 158: ......
Page 159: ......
Page 160: ......
Page 161: ......
Page 162: ......
Page 163: ......
Page 164: ......
Page 165: ......
Page 166: ......
Page 167: ......
Page 168: ......
Page 169: ......
Page 170: ......
Page 171: ......
Page 172: ......
Page 173: ......
Page 174: ......
Page 175: ......
Page 220: ......
Page 221: ......
Page 222: ......
Page 223: ......
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 237: ...Version 04 2004 50078754 Accessories for CO2 incubators Illustrations and describtion...
Page 249: ...Footprint HERAcell...
Page 250: ...Footprint HERAcell...