4.14.4 ADJUSTMENTS - O
2
METERING AND REGULATION CIRCUIT
In general, a comparative measurement with O
2
gas addition to the recommended
values for the individual sensor should be performed.
Comparative measurement: O
2
regulation circuit
Perform the comparative measurements at the following O
2
concentrations:
Sensor measurement range 1...21% O
2
: approx. 5% O
2
Sensor measurement range 5...90% O
2
: approx. 90% O
2
Give the unit sufficient time to stabilize before reading the measured values:
-
For test gas addition:
No sooner than 45 min. after all regulating circuits
have reached equilibrium
-
For units at operating temp.,
with gas, stable humidity:
The comparative measurement can be made im-
mediately.
Performing the comparative measure-
ment:
(O
2
reference measuring instrument)
Extract three gas samples through the metering opening in the gas diaphragm/glass
door.
Read and average the values, then compare with the displayed value.
Sensor measurement range 1...21% O
2
Deviation:
< ± 0.5 % O
2
No adjustment necessary.
> ± 0.5 % O
2
Perform the adjustment (see below)
Sensor measurement range 5...90% O
2
Deviation:
< ± 2 % O
2
No adjustment necessary.
> ± 2 % O
2
Perform the adjustment (see below)
O
2
adjustment procedure:
Press and hold down the -cal- , -i- , and -O
2
- keys simultaneously until the O
2
dis-
play flashes.
Set the reference measured value.
Press any key to return to the normal operating mode.
Summary of Contents for HERAcell 150
Page 82: ...CO2 Incubator HERAcell 150 Operating Instructions 82 16 Certificate of decontamination...
Page 83: ...CO2 Incubator HERAcell 150 Operating Instructions 83 16 Certificate of decontamination...
Page 84: ...Internet http www kendro com...
Page 85: ......
Page 87: ...Valid 10 2005 50047292 B Gas Cylinder Change Over Unit GM 2 Operating instructions GB...
Page 106: ......
Page 152: ......
Page 153: ......
Page 154: ......
Page 155: ......
Page 156: ......
Page 157: ......
Page 158: ......
Page 159: ......
Page 160: ......
Page 161: ......
Page 162: ......
Page 163: ......
Page 164: ......
Page 165: ......
Page 166: ......
Page 167: ......
Page 168: ......
Page 169: ......
Page 170: ......
Page 171: ......
Page 172: ......
Page 173: ......
Page 174: ......
Page 175: ......
Page 220: ......
Page 221: ......
Page 222: ......
Page 223: ......
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 237: ...Version 04 2004 50078754 Accessories for CO2 incubators Illustrations and describtion...
Page 249: ...Footprint HERAcell...
Page 250: ...Footprint HERAcell...