CO
2
Incubator HERAcell
®
150 Operating Instructions
22
4.
Description of the device
4.3
Safety devices
The device is equipped with the following safety devices:
• A door switch interrupts the CO
2
and O
2
supply and the work space heating
when the glass door is opened.
• The optional gas guard switches the gas supply over to a full gas cylinder.
• An independent thermal protection protects the samples from harmful over-
heating in case of failures.
• A pressure compensation opening ensures pressure compensation in the
device work space.
• Audible and visual alarms indicate failures during operation.
4.4
Work space atmosphere
In the work space of the incubator, the particular physiological ambient condi-
tions for the preparation and cultivation of cell and tissue cultures are simula-
ted. The work space atmosphere is determined by the following factors:
• Temperature,
• Relative humidity,
• CO
2
concentration,
• O
2
concentration (option).
Temperature:
To ensure undisturbed operation, the temperature in the operating room must
be at least 18° C and the incubation temperature must be at least 3° C higher
than this temperature.
The heating system controls the incubation temperature from this temperature
threshold up to 55° C. The principle of air jacket heating and the additional, se-
parate heating of the outer door and glass door/gas tight screen minimize the
risk of condensate formation at the side walls, at the ceiling of the work space,
and at the glass door/gas tight screen.
Relative humidity:
The water tray of the work space can hold 3.0 liters of processed water. The hea-
ting of the work space supports the condensation of the water, thereby ensu-
ring a constant humidity within the work space. Under normal operating conditions
and at the usual incubation temperature of 37° C, a constant relative humidity
of approx 95 % is achieved in the work space.
A high/low switch-over function allows the selection of two humidity settings:
• Select
setting 0 (High)
for a relative humidity of
approx 95 %
(standard setting).
• Select
setting I (Low)
for a relative humidity of
approx 90 %
.
If heated containers are removed and put back into the work space, the elevated
humidity and the cooling may cause condensate to form on the outer sides of
the container.
The low humidity regulation effectively prevents the formation of conden-
sate on containers.
Summary of Contents for HERAcell 150
Page 82: ...CO2 Incubator HERAcell 150 Operating Instructions 82 16 Certificate of decontamination...
Page 83: ...CO2 Incubator HERAcell 150 Operating Instructions 83 16 Certificate of decontamination...
Page 84: ...Internet http www kendro com...
Page 85: ......
Page 87: ...Valid 10 2005 50047292 B Gas Cylinder Change Over Unit GM 2 Operating instructions GB...
Page 106: ......
Page 152: ......
Page 153: ......
Page 154: ......
Page 155: ......
Page 156: ......
Page 157: ......
Page 158: ......
Page 159: ......
Page 160: ......
Page 161: ......
Page 162: ......
Page 163: ......
Page 164: ......
Page 165: ......
Page 166: ......
Page 167: ......
Page 168: ......
Page 169: ......
Page 170: ......
Page 171: ......
Page 172: ......
Page 173: ......
Page 174: ......
Page 175: ......
Page 220: ......
Page 221: ......
Page 222: ......
Page 223: ......
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 237: ...Version 04 2004 50078754 Accessories for CO2 incubators Illustrations and describtion...
Page 249: ...Footprint HERAcell...
Page 250: ...Footprint HERAcell...