Problem:
Possible cause:
Corrective action:
Processor version P002 is
shown as 000.
Unit beeps after it is
switched on.
The unit display shows [ = =
= ] after 1 minute.
Fan wheel not turning or
turning very slowly.
CO
2
tube and/or metering
cable wires are lying
against the fan motor.
Disassemble the
measuring cell. Check
the position of the CO
2
tube in the guide groove.
When reinstalling the
cell, make sure the cell’s
wiring is not obstructed
in the sample chamber
ceiling opening.
Western connector not
properly engaged.
Unplug the connector,
then plug it in
again.Engagement must
be audible.
Summary of Contents for HERAcell 150
Page 82: ...CO2 Incubator HERAcell 150 Operating Instructions 82 16 Certificate of decontamination...
Page 83: ...CO2 Incubator HERAcell 150 Operating Instructions 83 16 Certificate of decontamination...
Page 84: ...Internet http www kendro com...
Page 85: ......
Page 87: ...Valid 10 2005 50047292 B Gas Cylinder Change Over Unit GM 2 Operating instructions GB...
Page 106: ......
Page 152: ......
Page 153: ......
Page 154: ......
Page 155: ......
Page 156: ......
Page 157: ......
Page 158: ......
Page 159: ......
Page 160: ......
Page 161: ......
Page 162: ......
Page 163: ......
Page 164: ......
Page 165: ......
Page 166: ......
Page 167: ......
Page 168: ......
Page 169: ......
Page 170: ......
Page 171: ......
Page 172: ......
Page 173: ......
Page 174: ......
Page 175: ......
Page 220: ......
Page 221: ......
Page 222: ......
Page 223: ......
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 237: ...Version 04 2004 50078754 Accessories for CO2 incubators Illustrations and describtion...
Page 249: ...Footprint HERAcell...
Page 250: ...Footprint HERAcell...