4.14.3 PERFORMING ADJUSTMENTS CO
2
METERING AND REGULATION
CIRCUIT
Performing adjustments:
The following steps can be performed for units equipped with either a thermal conduc-
tivity or an infrared detector.
In general, a comparative measurement should be made, unless the actual value is
known, e.g., CO
2
display deviation at 0.0 % CO
2
.
Adjustment routine activation: Refer to the example in the operating instructions
(Chapter 6: Switch Function, Manual Zero-point Adjustment of the CO
2
Metering Sys-
tem)
Comparative meas-
urement:
CO
2
regulating circuit
Perform the comparative measurement at the user's normal CO
2
concentration, e.g.,
5.0 %. Make sure the unit has adequate time to reach equilibrium before taking a
reading:
-
For test addition: No sooner than 45 min. after all regulating circuits have reached
equilibrium.
-
For units at oper-
ating temp., with
gas, stable hu-
midity:
The measurement can be made immediately.
Comparative measurement proce-
dure:
(Portable IR meter or hand pump w. sample
tube 0..10 % CO
2
)
Extract three gas samples through the metering opening in the gas dia-
phragm/glass door.
Read and average the values, then compare with the displayed value.
Deviation:
<
0.5 % CO
2
No adjustment necessary.
>
0.5 % CO
2
Refer to Section 10.6 of the operating instruc-
tions.
Summary of Contents for HERAcell 150
Page 82: ...CO2 Incubator HERAcell 150 Operating Instructions 82 16 Certificate of decontamination...
Page 83: ...CO2 Incubator HERAcell 150 Operating Instructions 83 16 Certificate of decontamination...
Page 84: ...Internet http www kendro com...
Page 85: ......
Page 87: ...Valid 10 2005 50047292 B Gas Cylinder Change Over Unit GM 2 Operating instructions GB...
Page 106: ......
Page 152: ......
Page 153: ......
Page 154: ......
Page 155: ......
Page 156: ......
Page 157: ......
Page 158: ......
Page 159: ......
Page 160: ......
Page 161: ......
Page 162: ......
Page 163: ......
Page 164: ......
Page 165: ......
Page 166: ......
Page 167: ......
Page 168: ......
Page 169: ......
Page 170: ......
Page 171: ......
Page 172: ......
Page 173: ......
Page 174: ......
Page 175: ......
Page 220: ......
Page 221: ......
Page 222: ......
Page 223: ......
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 237: ...Version 04 2004 50078754 Accessories for CO2 incubators Illustrations and describtion...
Page 249: ...Footprint HERAcell...
Page 250: ...Footprint HERAcell...