Incubator HERAcell
®
150 Operating Instructions
57
8.
Shut-down
8.1
Shutting the device down
CAUTION! – Contamination hazard!
If the work space surfaces are contaminated,
germs my be transferred to the environment
of the device.
In case of a shut-down, the device must be
decontaminated!
1. Remove culture containers and all accessories from the work space.
2. Pump water off (see Section 9.2).
3. Fill in 350 ml of fresh processed water and start the ContraCon decontami-
nation routine.
After the ContraCon decontamination routine has been run, disconnect the device
from the power and gas supply:
4. Wipe device dry.
5. Turn device off using the power switch.
6. Unplug power connector and protect it against accidental reconnection.
7. Close the CO
2
/ O
2
/N
2
supply system shut-off valves.
8. Disconnect gas pressure hoses from sleeve at the rear of the device.
9. Until the device is shut down, the work space must be continuously venti-
lated: Leave the glass door and the outer door open and secure them in this
state.
Summary of Contents for HERAcell 150
Page 82: ...CO2 Incubator HERAcell 150 Operating Instructions 82 16 Certificate of decontamination...
Page 83: ...CO2 Incubator HERAcell 150 Operating Instructions 83 16 Certificate of decontamination...
Page 84: ...Internet http www kendro com...
Page 85: ......
Page 87: ...Valid 10 2005 50047292 B Gas Cylinder Change Over Unit GM 2 Operating instructions GB...
Page 106: ......
Page 152: ......
Page 153: ......
Page 154: ......
Page 155: ......
Page 156: ......
Page 157: ......
Page 158: ......
Page 159: ......
Page 160: ......
Page 161: ......
Page 162: ......
Page 163: ......
Page 164: ......
Page 165: ......
Page 166: ......
Page 167: ......
Page 168: ......
Page 169: ......
Page 170: ......
Page 171: ......
Page 172: ......
Page 173: ......
Page 174: ......
Page 175: ......
Page 220: ......
Page 221: ......
Page 222: ......
Page 223: ......
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 237: ...Version 04 2004 50078754 Accessories for CO2 incubators Illustrations and describtion...
Page 249: ...Footprint HERAcell...
Page 250: ...Footprint HERAcell...