150 Operating Instructions
68
10.
Maintenance
10.4
Temperature calibration procedure
1. Activate calibration:
Keep the
key depressed for 5 seconds.
All operating panel indicators flash.
2. Display the nominal value:
Press the
key.
The preset value of 37° C is displayed.
3. Enter the measured value (destination value):
Press the
+
keys.
or
Press the
+
keys.
Destination value e.g. 36.4° C
4. Accept the destination value:
Press the
key.
The temperature display momentarily shows “CAL”,
then the corrected actual value (measured destination value 36.4° C)
is displayed.
5. Cancel the calibration process:
Press any key.
The temperature display and the CO
2
display show the actual values.
Summary of Contents for HERAcell 150
Page 82: ...CO2 Incubator HERAcell 150 Operating Instructions 82 16 Certificate of decontamination...
Page 83: ...CO2 Incubator HERAcell 150 Operating Instructions 83 16 Certificate of decontamination...
Page 84: ...Internet http www kendro com...
Page 85: ......
Page 87: ...Valid 10 2005 50047292 B Gas Cylinder Change Over Unit GM 2 Operating instructions GB...
Page 106: ......
Page 152: ......
Page 153: ......
Page 154: ......
Page 155: ......
Page 156: ......
Page 157: ......
Page 158: ......
Page 159: ......
Page 160: ......
Page 161: ......
Page 162: ......
Page 163: ......
Page 164: ......
Page 165: ......
Page 166: ......
Page 167: ......
Page 168: ......
Page 169: ......
Page 170: ......
Page 171: ......
Page 172: ......
Page 173: ......
Page 174: ......
Page 175: ......
Page 220: ......
Page 221: ......
Page 222: ......
Page 223: ......
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 237: ...Version 04 2004 50078754 Accessories for CO2 incubators Illustrations and describtion...
Page 249: ...Footprint HERAcell...
Page 250: ...Footprint HERAcell...